(Created page with "==Section 4 - CARDIAC== ===4.14 VENTRICULAR ASSIST DEVICES=== This Parameter applies to the management of patients who have a ventricular assist device (VAD) also referred to...") |
|||
Line 8: | Line 8: | ||
** Provide OXYGEN @ 100% via NRB mask or assist ventilations/oxygenation as indicated. | ** Provide OXYGEN @ 100% via NRB mask or assist ventilations/oxygenation as indicated. | ||
* Listen to heart sounds. The heart sounds may be faint. | * Listen to heart sounds. The heart sounds may be faint. | ||
* Check for a functioning VAD: These devices are usually implanted below the heart in the abdominal cavity area. Use a stethoscope to listen for constant humming or whirling sound. This sound confirms the VAD is functioning. | * '''Check for a functioning VAD''': These devices are usually implanted below the heart in the abdominal cavity area. Use a stethoscope to listen for constant humming or whirling sound. This sound confirms the VAD is functioning. | ||
** The nature of the VAD function provides positive constant flow (it is a centrifugal pump) therefore a pulse may not be palpable in these patients. | ** The nature of the VAD function provides positive constant flow (it is a centrifugal pump) therefore '''a pulse may not be palpable''' in these patients. | ||
* Blood Pressure: A manual blood pressure may not be obtainable. An automated blood pressure monitoring device may provide a reading with a narrow pulse pressure. | * '''Blood Pressure''': A manual blood pressure may not be obtainable. An automated blood pressure monitoring device may provide a reading with a narrow pulse pressure. | ||
** Treatment of these patients must be based on the mean arterial pressure (MAP). The normal MAP should be between 60 and 90 mm/Hg. | ** Treatment of these patients must be based on the mean arterial pressure (MAP). The normal MAP should be between 60 and 90 mm/Hg. | ||
* Pulse Oximetry: These readings may not be accurate due to the continuous flow provided by the VAD. | * '''Pulse Oximetry''': These readings may not be accurate due to the continuous flow provided by the VAD. | ||
** Monitor side capnography for ventilation and oxygenation effectiveness. | ** Monitor side capnography for ventilation and oxygenation effectiveness. | ||
Line 19: | Line 19: | ||
* If the patient is unresponsive to stimuli: | * If the patient is unresponsive to stimuli: | ||
** Open the airway and ventilate/oxygenate as necessary. | ** Open the airway and ventilate/oxygenate as necessary. | ||
** Obtain a blood glucose level and treat per 5.10 Hypoglycemia parameter as needed. | ** Obtain a blood glucose level and treat per [[Hypo or Hyperglycemia|5.10 Hypoglycemia]] parameter as needed. | ||
* Listen for a functioning VAD by auscultating over the abdomen. | * Listen for a functioning VAD by auscultating over the abdomen. | ||
** If the VAD has a whirling sound it is properly functioning. | ** If the VAD has a whirling sound it is properly functioning. | ||
** DO NOT PERFORM CPR! | ** '''DO NOT PERFORM CPR!''' | ||
*** A VAD or LVAD is surgically implanted to the left ventricle. External cardiac compressions may dislodge the VAD causing death. | *** A VAD or LVAD is surgically implanted to the left ventricle. External cardiac compressions may dislodge the VAD causing death. | ||
* If the VAD is silent have the caregiver or family member check the device external pump controller and power supply. | * If the VAD is silent have the caregiver or family member check the device external pump controller and power supply. | ||
** LISTEN TO THE CAREGIVER! These individuals are experts in these devices and know/understand how to operate and troubleshoot the system. | ** '''LISTEN TO THE CAREGIVER!''' These individuals are experts in these devices and know/understand how to operate and troubleshoot the system. | ||
** VAD patients are instructed to always have a backup bag with 2 charged batteries and a second pump controller. | ** VAD patients are instructed to always have a backup bag with 2 charged batteries and a second pump controller. | ||
** Contact the LVAD coordinator on call (based at Florida South). | ** Contact the LVAD coordinator on call (based at Florida South). | ||
Line 41: | Line 41: | ||
'''TRANSPORT OF PATIENT:''' | '''TRANSPORT OF PATIENT:''' | ||
* ALL VAD patients must be transported to the VAD center. | * '''ALL VAD patients must be transported to the VAD center'''. | ||
** Currently Florida Hospital Orlando is the only local approved VAD center. | ** Currently '''Florida Hospital Orlando''' is the only local approved VAD center. | ||
** Contact the VAD coordinator as soon as possible and notify them of the transport. | ** Contact the VAD coordinator as soon as possible and notify them of the transport. | ||
* Don’t forget the family member/caregiver! | * Don’t forget the family member/caregiver! | ||
** The family member/caregiver must ACCOMPANY the patient in the ambulance. | ** The family member/caregiver must '''ACCOMPANY''' the patient in the ambulance. | ||
** These individuals have intimate knowledge of the patient’s condition, the system and usually have legal authority to sign on behalf of the patient. | ** These individuals have intimate knowledge of the patient’s condition, the system and usually have legal authority to sign on behalf of the patient. | ||
* Bring all VAD equipment, backup bags, charger and batteries with the patient! | * Bring all VAD equipment, backup bags, charger and batteries with the patient! |
Revision as of 19:34, 15 March 2013
Section 4 - CARDIAC
4.14 VENTRICULAR ASSIST DEVICES
This Parameter applies to the management of patients who have a ventricular assist device (VAD) also referred to a left ventricular assist device (LVAD) implanted. The VAD is a mechanical pump used to support the heart’s pumping function and positive blood flow on individuals who have severe heart failure.
- Perform initial assessment: vital signs, oxygen saturation, pre-treatment 12-Lead ECG, focused history (SAMPLE, OPQRST), and physical exam. Inquire if the patient has an old ECG.
- Provide OXYGEN @ 100% via NRB mask or assist ventilations/oxygenation as indicated.
- Listen to heart sounds. The heart sounds may be faint.
- Check for a functioning VAD: These devices are usually implanted below the heart in the abdominal cavity area. Use a stethoscope to listen for constant humming or whirling sound. This sound confirms the VAD is functioning.
- The nature of the VAD function provides positive constant flow (it is a centrifugal pump) therefore a pulse may not be palpable in these patients.
- Blood Pressure: A manual blood pressure may not be obtainable. An automated blood pressure monitoring device may provide a reading with a narrow pulse pressure.
- Treatment of these patients must be based on the mean arterial pressure (MAP). The normal MAP should be between 60 and 90 mm/Hg.
- Pulse Oximetry: These readings may not be accurate due to the continuous flow provided by the VAD.
- Monitor side capnography for ventilation and oxygenation effectiveness.
UNRESPONSIVE PATIENT
- If the patient is unresponsive to stimuli:
- Open the airway and ventilate/oxygenate as necessary.
- Obtain a blood glucose level and treat per 5.10 Hypoglycemia parameter as needed.
- Listen for a functioning VAD by auscultating over the abdomen.
- If the VAD has a whirling sound it is properly functioning.
- DO NOT PERFORM CPR!
- A VAD or LVAD is surgically implanted to the left ventricle. External cardiac compressions may dislodge the VAD causing death.
- If the VAD is silent have the caregiver or family member check the device external pump controller and power supply.
- LISTEN TO THE CAREGIVER! These individuals are experts in these devices and know/understand how to operate and troubleshoot the system.
- VAD patients are instructed to always have a backup bag with 2 charged batteries and a second pump controller.
- Contact the LVAD coordinator on call (based at Florida South).
- If the VAD is inoperative despite troubleshooting and battery replacement, then CPR is indicated as per cardiac arrest parameter.
RESPONSIVE PATIENT:
- Perform a subjective interview to determine chief complaint and history of present illness.
- Monitor ECG and obtain a 12 lead ECG if chest pain, ischemic symptoms, shortness of breath or weakness is reported.
- Any cardiac dysrhythmias must be treated per the appropriate practice parameter.
- If electrical therapy is indicated on a conscious patient, the patient may be sedated (time permitting) as per parameter 2.04 Analgesia/Sedation.
- Establish IV access and provide fluid resuscitation as indicated based on patient’s condition and hydration status.
- Assess for any other concurrent medical conditions or symptoms and treat per the appropriate parameter.
TRANSPORT OF PATIENT:
- ALL VAD patients must be transported to the VAD center.
- Currently Florida Hospital Orlando is the only local approved VAD center.
- Contact the VAD coordinator as soon as possible and notify them of the transport.
- Don’t forget the family member/caregiver!
- The family member/caregiver must ACCOMPANY the patient in the ambulance.
- These individuals have intimate knowledge of the patient’s condition, the system and usually have legal authority to sign on behalf of the patient.
- Bring all VAD equipment, backup bags, charger and batteries with the patient!